SH/T 0068-2002 Determination of density or relative density of engine coolants and their concentrates (hydrometer method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English name / replacement relationship

2. Applicable scope and non-applicability

3. Method principle

The original text does not provide an independent "principle" section. The method is the hydrometer method: keep the sample at constant temperature at 20℃±0.2℃, place the hydrometer into the sample, and after it becomes stationary and suspended, read the hydrometer reading (for transparent samples, read the lower edge of the liquid surface; for turbid samples, read the upper edge above the main liquid surface and correct it).

4. Reagents and materials (purity / preparation)

5. Apparatus and equipment

Table 1 Hydrometer specifications

Item number Range density/(kg/m³) Relative density
1 1000–1050 1.000–1.050
2 1050–1100 1.050–1.100
3 1100–1150 1.100–1.150
4 1150–1200 1.150–1.200
5 1200–1250 1.200–1.250
6 1250–1300 1.250–1.300
7 1300–1350 1.300–1.350

Appendix A (normative appendix) Thermometer specifications

Table A1 Thermometer specifications

Item Index
Range/℃ -20~102 (the original text is "yi20 yi102", requires confirmation against the source standard (source-text extraction))
Immersion depth Full immersion
Minimum division value/℃ 0.2
Long-line division value/℃ 1
Numerical scale/℃ 2
Scale error/℃ Not more than 0.15 (the original text is "0.qiao", requires confirmation against the source standard (source-text extraction))
Expansion chamber allowed to be heated to/℃ 150
Overall length/mm 415–425
Rod diameter/mm 6.0–8.0
Mercury bulb length/mm 15~20 (the original text is "qiao~20", requires confirmation against the source standard (source-text extraction))
Mercury bulb outer diameter Not greater than the outer diameter of the rod
Distance from the bottom of the mercury bulb to -20℃/mm 35–50
Total length of scale range/mm 305–350

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analytical procedure

9. Qualitative and quantitative determination and calculation formulas

10. Precision

Judge the reliability of test results according to the following provisions (95% confidence level).

11. Report

Not provided in the available source.

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A

Question Key answer
Standard number and year? SH/T 0068-2002
Which standard does it replace? SH/T 0068-91
Which foreign standard is equivalently adopted? ASTM D 1122-97a
Method name? Determination of density or relative density of engine coolants and their concentrates (hydrometer method)
Scope of application? Determination of density or relative density of engine coolants and their concentrates using a hydrometer
Test temperature? 20℃±0.2℃
One of the main differences from ASTM (temperature)? The original text 15.5℃±0.3℃ was revised to 20℃±0.2℃
Minimum scale of the hydrometer? 0.5 kg/m³ (the value and unit are unclear in the source transcription and remain unverified)
Change in hydrometer measurement range? Expanded from 40 kg/m³ to 50 kg/m³
Number of hydrometers? 7 pieces (Table 1)
Requirement for the inner diameter of the hydrometer cylinder? At least 25.4 mm larger than the outer diameter of the hydrometer
Requirement for the cylinder height? The sample height must be at least 25.4 mm more than the portion of the hydrometer immersed below the liquid surface
Thermometer requirement? Complies with Appendix A, or an ordinary thermometer of 0℃~100℃, with a scale division of 0.2℃ or 0.1℃
Water bath temperature? 20℃±0.2℃
Sampling reference standard? SH/T 0065
Treatment of stratified sample? Pour into a 500 mL separatory funnel, let stand for not less than 20℃ 3 h, and separate the lower layer sample
Sample temperature before pouring into the cylinder? About 18℃
Reading method for transparent samples? Read the lower edge of the meniscus (the straight line where the liquid surface is tangent to the scale)
Reading method for cloudy samples? Read the upper edge above the main liquid surface, and correct using a value at the same height
How is the correction value determined? Observe the height above the main liquid surface using a transparent solution with the same surface tension as the sample (such as ethylene glycol or propylene glycol solution)
Repeatability limit? In the density range 1000.0~1130.0 kg/m³, not greater than 1.2 kg/m³ (relative density 0.0012)
Reproducibility limit? In the density range 1000.0~1130.0 kg/m³, not greater than 1.2 kg/m³ (relative density 0.0012)
Definition of density? The mass of a substance per unit volume at a specified temperature, kg/m³
Definition of relative density? The ratio of the density of a substance at a specified temperature to the density of water at the same temperature
Use of density/relative density? To roughly determine freezing point, boiling point, and glycol content; used as a production control indicator
What is Appendix A? An appendix of the standard, thermometer specifications
First publication date? 1991 year
Publication/implementation date? Published on 2002-05-31, implemented on 2002-07-01

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.