SH/T 0793-2007 Determination of aromatic carbon content in hydrocarbon oils (High-resolution nuclear magnetic resonance method)

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full title / English title / replacement relationship

2. Scope of application and non-applicability

3. Method principle

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

Table 1 Instrument conditions for ¹H NMR samples and continuous-wave (CW) measurement:

Table 2 Instrument conditions for ¹H and ¹³C NMR spectra samples and pulsed Fourier transform measurement:

7. Sampling and test specimen preparation

8. Analytical procedure

Procedure A—— uses a continuous wave (CW) NMR spectrometer to determine ¹H NMR:

Procedure B—— uses a pulsed Fourier transform (FT) NMR spectrometer to determine ¹H NMR:

Procedure C—— uses a pulsed Fourier transform (FT) NMR spectrometer to determine ¹³C NMR:

9. Qualitative and Quantitative Determination and Calculation Formulas

10. Precision

11. Report

12. Referenced Documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Reference for Questions and Answers

Question Key points of answer
What does this standard measure? Aromatic hydrogen content in petroleum products (steps A, B) and aromatic carbon content (steps C), SH/T 0793—2007
Which samples are applicable? Kerosene, heavy distillate oils, mineral oils, lubricating oils, liquids derived from coal, and other fractions that are completely soluble in chloroform at room temperature
What is not applicable? Samples containing more than 1% (mass fraction) olefins or phenolic compounds; aromatic hydrocarbon mass fraction is not measured
Detection lower limit? FT spectrometer: 0.1% (mole fraction) aromatic hydrogen, 0.5% (mole fraction) aromatic carbon; CW spectrometer: 0.5% (mole fraction) aromatic hydrogen
What internal standard is used? TMS, both ¹H and ¹³C spectral chemical shifts are assigned to 0.0ppm
TMS concentration requirement? For ¹H spectrum measurement, use deuterated chloroform containing TMS 0.03% (volume fraction); for ¹³C spectrum measurement, use deuterated chloroform containing TMS 1% (volume fraction)
Deuterated chloroform requirements? Degree of deuteration not less than 99.8%; for ¹H NMR, residual chloroform in deuterated chloroform must be less than 0.2% (volume fraction)
What is the relaxation reagent? 2,4- chromium(III) pentanedionate (chromium acetylacetonate), recommended for use only with ¹³C NMR, concentration 20mmol/L (approximately 10mg/mL), weigh 10mg per milliliter of final solution
Aromatic hydrogen integration range? 5.0ppm–10.0ppm (A region); aliphatic hydrogen region -0.5ppm–5ppm (B region)
Aromatic carbon integration range? 100ppm–170ppm (A region); aliphatic carbon region -10ppm–70ppm (B region)
Calculation formula? Aromatic hydrogen or aromatic carbon content = [A/(A+B)]×100%
Result unit and retention? Mole fraction of aromatic hydrogen atoms and/or mole fraction of aromatic carbon atoms, retained to one decimal place
Repeatability? Aromatic hydrogen 0.32X; aromatic carbon 0.59X (requires confirmation against the source standard (source-text extraction)), X is the average of two results
Reproducibility? Aromatic hydrogen 0.49X (requires confirmation against the source standard (source-text extraction)); aromatic carbon 1.37X (requires confirmation against the source standard (source-text extraction)), X is the average of two results
Minimum sample amount? Approximately 10mL, should be sufficient to allow two determinations
How are waxy samples handled? Heat to approximately 60℃, mix thoroughly with a high-shear mixer before sampling, and if necessary transfer with a pipette preheated to approximately 60℃
Which documents are cited? GB/T 4756, SH/T 0729, ASTM E386, ASTM D5292-99 (2004)
Modified adoption of source standard? ASTM D5292-99 (2004), main differences are the TMS solution preparation requirements and the rewriting of repeatability and reproducibility statements, and deletion of the keywords section

Related public products

Method references do not confirm that a product is suitable for this edition. Check its certificate and availability.

Search related products

Source and review status

This translation has not received a complete independent technical review. Missing sections and unclear source values remain identified above.