SH/T 0090-1991 Engine coolant freezing point determination method

AI-translated method card — human verification pending. This is not an official translation of the standard. Unclear source values remain unverified; check the original standard and product certificate before use.

Card type: Test method

1. Method number / full name / English name / replacement relationship

2. Scope of application and non-applicability

3. Method principle

Take a test sample of 75~100 mL, inject it into a cooling tube with a stirring device, cool the sample with a refrigerant or other method, record the temperature of the sample every minute, and draw a time-temperature curve. The projection point on the vertical axis of the intersection of the cooling curve and the crystallization curve is the freezing point of the sample. If the sample exhibits supercooling, then the highest temperature reached when the temperature rises after supercooling is the freezing point of the sample. (Chapter 4)

4. Reagents and materials (purity / preparation)

5. Instruments and equipment

6. Recommended operating conditions

7. Sampling and sample preparation

8. Analysis steps

9. Qualitative and quantitative determination and calculation formula

10. Precision

11. Report

12. Referenced documents

13. Source and verification

This page is an AI translation of the available Chinese method card. Human verification is pending. It is not an official translation of the standard. Check the source standard and product certificate before use.

14. Quick Q&A reference

Question Key answer points
What does this standard apply to? Engine coolant and its concentrated liquid (Chapter 1)
What is the definition of freezing point? In the absence of supercooling, the temperature at which the coolant begins to crystallize; in the presence of supercooling, the highest temperature reached when it rapidly rises after initially forming crystals (Chapter 3)
How much sample is measured? 75~100 mL (Chapter 4, 7.1)
What is the stirring speed? 60~80 times per minute (7.2)
What is the cooling rate controlled to? Within 1 ℃/min (the source says "19C/min", requires confirmation against the source standard (source-text extraction)) (7.3)
What is the dry ice layer thickness requirement? At least 13 mm (the source says "13r nme", requires confirmation against the source standard (source-text extraction)) (6.3, 7.3)
How often is it recorded near the freezing point? Record once every 15 s (7.4)
When is crystallization induced? When the sample temperature reaches 0.5~1 ℃ above the expected freezing point (7.4)
Above what supercooling temperature is the test invalid? Above 1 ℃ (the source says "1T", requires confirmation against the source standard (source-text extraction)) (7.5 note)
How long must observation continue after crystallization? At least 10 min (7.5)
How is the freezing point determined? The projection point on the vertical axis of the intersection of the cooling curve and the crystallization curve; when supercooled, it is the highest temperature reached on recovery (7.7)
Repeatability requirement? The difference between the two results and the arithmetic mean shall not exceed 0.3 ℃ (original text "0.39C", requires confirmation against the source standard (source-text extraction)) (8.1)
Reproducibility requirement? The difference between results from different laboratories and the arithmetic mean shall not exceed 0.6 ℃ (original text "0.69 C", requires confirmation against the source standard (source-text extraction)) (8.2)
To what precision are results reported? 0.1 ℃ (original text "0.19C", requires confirmation against the source standard (source-text extraction)) (Chapter 9)
How is the concentrate handled? Dilute to the specified concentration according to SH/T 0065, then measure the freezing point (6.2)
What if two phases appear in the sample? Generally, separation is not necessary; the precipitate may be preservative, and removing it would affect the result (6.2)
Which foreign standard is referenced? ASTM D1177-88 (additional notes)
Where are the thermometer technical specifications? Annex A (supplementary document) Table A1